EqLifeBand

WEARABLE

A rugged wearable device that monitors vital signs in real time, ensuring safety for soldiers, construction workers, and high-risk professionals through continuous health tracking and timely alerts.

Project under

InventIndia Innovations, IN

Client

Equivital, UK

THE CHALLENGE

Heat stress causes thousands of workplace deaths every year, hitting construction, manufacturing, mining, and military training hardest. Existing vital-monitoring armbands (Kenzen, SlateSafety, Biovotion, Polar, Whoop) are uncomfortable for the long, physically demanding shifts these workers actually do.

400+ heat-related workplace deaths since 2011 - U.S. Bureau of Labor Statistics

The Process

START — PILOT BATCH

From defining the goals to the pilot batch, in order.

Explore

Prototype

Refine

Validate

Research

01

02

Concept Exploration

Product Architecture

03

04

POC — EVT

Form Development

05

06

Final Device Design

Material testing

Strap Design

07

08

Chargers Design

T-1 Samples

09

10

Pilot Batch - DVT

Design Goals

Five components had to work together as one system - the wearable, its strap, and the charging infrastructure behind it.

01

Vital Monitoring Device

Worn on the upper arm with PPG-based tracking, lightweight, discreet, and comfortable for long-term wear.

02

Detachable Arm Strap

Attaches and detaches easily from the body, breathable for comfort, built to ATEX material standards.

03

ATEX Compliant

Fully rated for hazardous environments with explosive gas or dust, with intrinsically safe components.

04

Single Dock Charger

A compact, precise charging dock that aligns the device automatically for effortless daily charging.

05

Multi-Dock Charger

Charges up to 25 devices at once ,organized, secure, and built for clinical or industrial settings.

Concept Exploration

Two attachment mechanisms were explored side by side, then the enclosure architecture was optimized around whichever proved sleekest to build.

Spring Release Mechanism

A spring-loaded plunger grips a lip in the cavity; a leaf-spring trigger releases the device with a single press.

Bayonet Mechanism

Flanges align and twist to lock firmly in place, with a reverse twist for easy removal and a positive end-rotation lock.

Product Architecture

Three product architectures were explored to test different internal volumes and integrate the two shortlisted attachment mechanisms - each pushed toward the sleekest possible enclosure. A dimension evaluation across the candidates landed the final build at 13.50mm thick, shaving nearly half a millimeter off the bulkiest option.

POC - EVT

A batch of 12 devices was made for EVT phase.

Hand injection molding was used to make 20 pieces for testing purposes. Injection molding was chosen to ensure sufficient strength for the bayonet mechanism. 3D printed becomes quite brittle.

Form development

Different packages were explored for volumetric breakdown. The main idea was to make the device look as sleek as possible, perceptually. The internal assembly across all the mockups was the same.

Final Device Design

06 - 10

After the testing phase of initial EVT prototypes, the final Industrial Design round was done, and this design concept was taken forward.

Strap Design

Seven clasping and perforation variants were cut and refined from the winning strap concept, then hand-built into wearable POC1 units in the workshop so the earliest device prototypes had something to strap on.

Strap, Tooling & Validation

Engineering and tooling support then carried the design through T0 and T1 sample rounds, coordinating directly with tooling vendors to validate fit and function, before a 60-piece pilot batch was cast from 3 vacuum-cast moulds for field testing.

In-house prototype

24-hour wear trial across 9 participants: 89% reported no discomfort from the module, and 100% never felt the need to switch arms

Prototypes at vendor’s

CNC Mockup

A CNC-machined dummy prototype was built to closely resemble the final product’s form and feel, giving visitors a hands-on experience at the American Industrial Hygiene Association (AIHA) exhibition booth.

Single dock charger ID

Multiple different types of architectures were built before coming down to this architecture with a base engulfing the device from all sides except top, to keep it safe from vibrations and physical movements. These are the 4 shortlisted variations of final industrial design tweaks.

Charging System

Two dedicated docks round out the system: a single-unit charger for everyday use, and a 25-slot dock for clinical or industrial fleets that need to charge and organize devices at scale.

Single Dock Charger

A compact, stable dock with a precise, secure fit - designed to make daily charging effortless.

Multi-Dock Charger

Charges up to 25 devices at once with organized, secure slots - compact yet robust for daily fleet use.

Each dock self-aligns magnetically - N52 magnets seated in the dock pull against metal pieces inside the device, so the pogo-pin contacts land correctly every time.

T-1 device samples

First batch was made for the top and bottom enclosure was made for part accuracy evaluation, as well as sink marks or flow marks. We observed a couple of things like the light guide fitment, bayonet mechanism working and ultrasound welding before going into the texture of the tool.

Sorry, please view it on desktop

Home

MORE PROJECTS

BACK TO WORK